STORE SEARCH CALIBRATING TEST SIEVES RECOMMENDED SIEVE RE-CERTIFICATION SCHEDULE

      Aperture                             Schedule*

  • 5 inches to .25 inches               80 uses or 36 months 
  • US No. 1/2 to US No. 12               70 uses or 30 months 
  • US No. 14 to US No. 40               60 uses or 24 months 
  • US No. 45 to US No. 140              50 uses or 18 months 
  • US No. 170 to US No. 325             35 uses or 12 months 
  • US No. 400 to US No. 635             20 uses or 6 months
sieves

MANUFACTURING DISTRIBUTION TECHNICAL SERVICES ​FOR ANALYTICAL INSTRUMENTATION

TELEPHONE (804) 318-3686 TOLL FREE  (866) 244-1578 "Without an active sieve re-certification program in place, you're simply guessing."  

Contact Us

Thank you for contacting us.
We will get back to you as soon as possible
Oops, there was an error sending your message.
Please try again later

RECALIBRATING TEST SIEVES Although a sieve may come from the factory with assurances of compliance, how long after use is it still considered calibrated? Depends on use

  • TYPE OF SIEVE MATERIAL
  • TYPE OF SIEVE SHAKER USED
  • NUMBER OF TESTS
  • SAMPLE CHARACTERISITICS

(Many states now require laboratories to calibrate their sieves every 6 months to remain certified)

test sieve sieve  calibrator

METHODS OF CALIBRATING SIEVES 1) MICROSCOPY-Primary Method-Directly traceable. Use of a caliper to calibrate      all or a minimum of 30 apertures on a sieve is labor intensive

2) DEDICATED IMAGE ANALYSIS    Dedicated image analysis has made huge strides recently, due to software development.

  NOTE: Recent innovations have brought this technology to the forefront. Costs have         been reduced  and operational macros have virtually eliminated operator training                See  Automatic Sieve Calibrator NEW AUTOMATIC SIEVE CALIBRATOR

QUICK COST RECOVERY! 2 MINUTE AUTOMATIC PROCEDURE ELIMINATES UP TO 90% LABOR

Caliper

test sieve

​3)  CALIBRATION STANDARDS

 WIDE DISTRIBUTION CRUSHED QUARTZ 

 Irregular shape of crushed quartz powder leads to confusion in exact    parameter required to pass sieve apertures.  These  glass microspheres  calibrate close to 80% of the sieving medium

sieve calibration

PRECISION SODA-LIME GLASS MICROSPHERES

are certified using electro-formed sieves, and measure the minimum  ​mean aperture size, (Dimension that retains spherical particles) They can only be used  once however, because some damage is presumed to occur during the calibration test.

NEW! ON SITE AND REMOTE SIEVE CALIBRATION SERVICE

4) PERFORMANCE TESTING USING A MASTER STACK OF SIEVES (ASTM E 2427)  NEW IN 2006!    Calibration Sieves may be used as a master sieve stack, to compare results to a working stack of sieves,                  using a representative sample.This can be a very cost effective method of calibrating working sieves.     Sample degradation can be a concern.

5 THIRD PARTY CERTIFICATION   Third party certification is always an option at a cost of about $90 to $150 per sieve, depending on the service required.   Available options are Sieve Inspection ASTM E-11 09, Sieve Certification ASTM E-11 04, Sieve Calibration ASTM E-11 09

WAYS TO EXTEND USABLE SIEVE LIFE

  • Use an ultrasonic sieve cleaner

WHITESTONE VA TEL (804) 318-3686 FAX (703) 991-7133 WWW.QCLABEQUIPMENT.COM Copyright 2012: QAQC Lab | Particle Measurement System | All rights reserved. 804-318-3686 |  sales@qclabequipment.com